Introduction
It is designed by the principle of infrared absorption spectroscopy for the determination of the concentration of carbon dioxide at the end of an exhaled breath,medical mainstream EtCO2 sensor module CM2200is widely used in monitoring the respiratory end tidal CO2 concentration and respiration rate.
It can be easily integrated to the monitors or anesthesia machines with an 8-pin socket and the preparation of the corresponding display/control software.
Features
1.EMI system can be against the interference of electromagnetic
2.Compatible with Respironics CapnoStat5 or CPT Protocols. TTL and R232 can be optional
3.Clinical waveform recognition ability is stronger
4.Low power consumption and small size
Specifications
Measurement range | CO2 Measurement Range A: 0~49mmHg, resolution is 0.1mmHg Measurement Range B: 50~150mmHg, resolution is 0.2mmHg |
CO2 accuracy | 0-40mmHg: ±2mmHg; 41-70mmHg:±5% of reading ; 71-100mmHg:±8% of reading; 100-150mmHg:±10% of reading |
Data interface | Highly configurable serial digital interface (TTL/RS232 Level) |
Response time | Detector: 28ms, System:200ms |
Respiration rate range | 2~150BPM |
Respiration rate accuracy | ±1 breath |
Data interface | TTL or RS232 |
Applications
1.High Flow Oxygen Therapy Instrument
2.Ventilator
3.Anesthesia Machine
*To get more product information, please contact info@gassensor.com.cn